1 Features
• Supports simultaneous and synchronized
acquisition of up to 24 PPG and a single-lead ECG
signal.
• Flexible allocation of 8 LED, 4 PD in each phase
• ECG signal chain:
– 1-lead ECG signal acquisition up to 2 kHz
– RLD output to drive a third electrode
– Input noise : 0.7 μV-rms
– Supports up to ±0.65 V differential DC offset
and +/-0.55 V common mode range
– Programmable INA gain: 11, 21
– High-pass filter with quick saturation recovery
– Integrated 300 Hz anti-aliasing low pass filter
– AC, DC lead-off detect: 2.9 nA to 92.5 nA
– Low power continuous lead-on detection
• Impedance signal chain
– Continuous low-power monitoring of impedance
between 2 electrodes
– 7 MΩ range
– 300 Ω noise on 620 kΩ
• PPG Transmitter:
– 8-Bit Programmable LED Current with range
adjustable from 25 mA to 250 mA
– Mode to fire two LEDs in parallel
– Support of 8 LEDs in Common Anode
configuration for SpO2, Multi-Wavelength HRM
• PPG Receiver:
– 2 parallel receivers (two sets of TIA/ filter)
– Supports 4 Time-Multiplexed Photodiode Inputs
to each Receiver
– 8-bit Ambient Offset subtraction at each TIA
input with 256 μA range
– 8-bit LED Offset Subtraction DAC with 64 μA
range
– Automatic ambient cancellation and dynamic
LED DC cancellation at TIA input
– Ambient rejection close to 100 dB up to 10 Hz
– Noise filtering with programmable bandwidth
– Trans-impedance Gain: 3.7 kΩ to 1 MΩ
• Accurate, Continuous Heart-Rate Monitoring:
– System SNR up to 109-dB at 16 μA PD current
– Low Current for Continuous Operation on a
Wearable Device with a Typical Value: 15-μA
for an LED, 20-μA for the Receiver
• External clock and internal oscillator modes
• Acquire data synchronized with system master
clock
• FIFO with 256-sample Depth
• SPI, I2C interfaces: Selectable by pin
• 2.6-mm × 2.5-mm DSBGA, 0.4-mm Pitch
• Supplies: Rx:1.7-1.9V (LDO Bypass); 1.9-3.6V
(LDO Enabled), Tx:3-5.5V, IO:1.7-RX_SUP
2 Applications
• Optical Heart-Rate Monitoring (HRM) for
Wearables, Hearables
• High performance ECG signal acquisition
• Synchronized PPG, ECG for BP estimation
• Heart-rate variability (HRV)
• Pulse oximetry (SpO2) measurements
3 Description
The AFE4950 device is an analog front-end (AFE)
for synchronized signal acquisition of PPG and ECG
signals. The device can also be used for optical biosensing
applications, such as heart-rate monitoring
(HRM) and saturation of peripheral capillary oxygen
(SpO2). The ECG signal chain supports 2- and 3-
electrode configurations and has an integrated Right
leg drive (RLD) buffer. The ECG signal chain can also
be used to do a continuous, low-power monitoring of
the impedance between pair of electrodes.
| 供应商 | 型号 | 品牌 | 批号 | 封装 | 库存 | 备注 | 价格 |
|---|---|---|---|---|---|---|---|
TI/德州仪器 |
23+ |
BGA36 |
5000 |
原厂授权代理,海外优势订货渠道。可提供大量库存,详 |
|||
TI(德州仪器) |
24+ |
BGA36 |
7350 |
现货供应,当天可交货!免费送样,原厂技术支持!!! |
|||
TI(德州仪器) |
24+ |
BGA36 |
1652 |
原装现货,免费供样,技术支持,原厂对接 |
|||
Texas Instruments |
25+ |
36-XFBGA DSBGA |
9350 |
独立分销商 公司只做原装 诚心经营 免费试样正品保证 |
|||
TI/德州仪器 |
25+ |
原厂封装 |
10280 |
原厂授权代理,专注军工、汽车、医疗、工业、新能源! |
|||
TI/德州仪器 |
25+ |
原厂封装 |
10280 |
||||
Texas Instruments |
36-DSBGA(2.63x2.46) |
60000 |
全新、原装 |
||||
TI/德州仪器 |
25+ |
原厂封装 |
11000 |
||||
TI/德州仪器 |
25+ |
原厂封装 |
10280 |
||||
TI德州仪器 |
22+ |
24000 |
原装正品现货,实单可谈,量大价优 |
AFE4950YBGR.A 资料下载更多...
AFE4950YBGR.A 芯片相关型号
Datasheet数据表PDF页码索引
- P1
- P2
- P3
- P4
- P5
- P6
- P7
- P8
- P9
- P10
- P11
- P12
- P13
- P14
- P15
- P16
- P17
- P18
- P19
- P20
- P21
- P22
- P23
- P24
- P25
- P26
- P27
- P28
- P29
- P30
- P31
- P32
- P33
- P34
- P35
- P36
- P37
- P38
- P39
- P40
- P41
- P42
- P43
- P44
- P45
- P46
- P47
- P48
- P49
- P50
- P51
- P52
- P53
- P54
- P55
- P56
- P57
- P58
- P59
- P60
- P61
- P62
- P63
- P64
- P65
- P66
- P67
- P68
- P69
- P70
- P71
- P72
- P73
- P74
- P75
- P76
- P77
- P78
- P79
- P80
- P81
- P82
- P83
- P84
- P85
- P86
- P87
- P88
- P89
- P90
- P91
- P92
- P93
- P94
- P95
- P96
- P97
- P98
- P99
- P100
- P101
- P102
- P103
- P104
- P105
- P106